Feature | Benefit |
DCCT based | Provides excellent stability and range linearity. |
Fast Reversal Rates | Reversal rates as fast as 4 seconds without affecting Bridge accuracy. |
Quick Measure Mode | First Measurement < 30 seconds |
New Mirco-Controller | Improved speed, functionality for the most efficient unit operation. |
National and Primary Lab Continuity | Used by leading NMI’s, Energy, Defense, Aeronautical and Space organizations. |
Up to 10,000 Amps | Can be expanded for shunt calibrations and DCCT’s. |
0.1 Ω to 100 kΩ | |||
Range (1:1 Ratio) | Accuracy (µΩ/Ω) | Range (1:10 Ratio) | Accuracy (µΩ/Ω) |
0.1 Ω to 0.1 Ω | 0.1 | 0.1 Ω to 1 Ω | 0.1 |
1 Ω to 1 Ω | 0.025 | 1 Ω to 10 Ω | 0.03 |
10 Ω to 10 Ω | 0.025 | 10 Ω to 100 Ω | 0.03 |
100 Ω to 100 Ω | 0.025 | 100 Ω to 1 kΩ | 0.03 |
1 kΩ to 1 kΩ | 0.025 | 1 kΩ to 10 kΩ | 0.03 |
10 kΩ to 10 kΩ | 0.1 | 1 kΩ to 12.9 kΩ | 0.03 |
10 kΩ to 100 kΩ | 0.05 |
10 µΩ to 1 Ω | ||
Range | Accuracy (µΩ/Ω) | With Range Extender (µΩ/Ω) |
10 µΩ to 100 µΩ | N/A | < 0.5 |
100 µΩ to 1 mΩ | N/A | < 0.4 |
1 mΩ to 10 mΩ | < 5 | < 0.3 |
10 mΩ to 100 mΩ | < 0.5 | < 0.2 |
100 mΩ to 1 Ω | < 0.1 | < 0.2 |