2020A Premium Power Analyzer

  • Updated replacement for 2010A wattmeter in the AccuLoss® Loss Measurement Systems
  • Basic power accuracy: < 0.005 %
  • Current ranges: 10
  • Voltage ranges: 2
  • Line-and-line and line-to-ground voltage measurements
  • Full power factor range
  • Complete waveform analysis
  • Total harmonic distortion analysis
  • Phasor analytical representation
Features & Benefits

The 2020A Power Analyzer is the latest development from Measurements International. It represents a new sampling technique for precise measurements of electrical power. Its diverse applications include R & D, product efficiency testing, transformer and reactor testing, and other power conversion applications.

This power analyzer has a basic power accuracy of < 0.005% as a self-contained standard, and a basic power accuracy of < 0.005 % as a self-contained standard. When combined with the MI series of 2500A High-Voltage Dividers and the MI series of 7020 Current Transformers, the 2020A Premium meets all the criteria, including instantaneous readings, speed, average and RMS voltages and waveform analysis, for calibrating both large and small reactors.

Specifications
Model No. 2020A Standard
Single-Phase
2020A Premium
Single-Phase
Input Channels 2 Voltage + 1 Current
Application Wattmeter Power Analyzer
Voltage
120 Volt Range Yes
Accuracy (ppm) ± 25
Linearity (ppm) ≤ 20
Input Impedance (Ω) 500 k
Frequency (Hz) 12 to 400
240 Volt Range Yes
Accuracy (ppm) ± 25
Linearity (ppm) ≤ 20
Input Impedance (Ω) 1 M
Frequency (Hz) 12 to 400
Current Measurement
Ranges (A) 0.005, 0.01, 0.02, 0.05, 0.1, 0.2, 0.5, 1, 2, 5 0.005, 0.01, 0.02, 0.05, 0.1, 0.2, 0.5, 1, 2, 5
Accuracy (ppm) ± 25 ± 25
Linearity (ppm) ≤ 20 ≤ 20
Input Impedance (Ω) ≤ 3.5 ≤ 3.5
Isolation (Vp/p) 600 600
Frequency (Hz) 12 to 400 12 to 400
Power Measurement
Power Factor 0 to > 0.5
Line-to-GND Accuracy (ppm) ± 25 to ± 50
Line-to-Line Accuracy (ppm) ± 35 to ± 75
Linearity (ppm) ≤ 20
Harmonic Distortions
Measurement
No Yes
Complete Waveform Analysis No Yes
Phasor Analysis No Yes
Operating Environment
Temperature (ºC) 15 to 40
Relative Humidity (%) 10 to 80
non-condensing
Line Voltage
Voltage (V) 100 to 240
Frequency (Hz) 47 to 63
Dimensions (mm) 175 × 440 × 380
Weight (kg) 15
Shipping Weight (kg) 20

Information Request