Featuring
The 1310B-X was designed to meet the requirements of automated DVM and calibrator calibration. Built around a low-thermal scanner and temperature controlled enclosure, the 1310B-X provides 11 high-stability standard reference resistors ranging from 0.1 Ω to 1 GΩ. The 1310B-X also allows users to add up to six additional resistor values to the unit. The system also includes two external channels that function as a connection hub for auxiliary AC and DC calibration equipment.
| Feature | Benefit |
| 11 resistors values ranging from 0.1 Ω up to 1 GΩ. With (6) Additional selectable values | Complete line of decade value standards in one temperature-controlled enclosure. |
| User-friendly touch screen. | Resistor tracking and history graph. |
| Internal resistance elements in a temperature controlled chamber. |
Excellent stability and extremely low- temperature coefficients. |
| Single output cable for direct plug-in. | Easy operation without the requirement for changing wires. |
| Calibrate calibrators and DVMs. | Best stability: < 2.5 μΩ/Ω/year. |
| Built-in 4-terminal scanner. | Combining multiple instruments (resistors, air bath, scanner) into one easy-to-use instrument. |
| Two external channels. | Connect to the resistance value of your choice. |
| Front-panel or GPIB-controlled. | Simplified operation. |
| Internally-mounted PT100 temperature sensor. | Connect to the front panel to monitor the internal oven. |
| Nominal Resistance (ohms) |
Nominal Resistance (± µΩ/Ω) (± ppm) |
24 Hour Stability (± µΩ/Ω) (± ppm) |
12 Month Stability (± µΩ/Ω) (± ppm) |
Temperature Coefficient (± µΩ/Ω/°C) (± ppm/°C) |
Max. Voltage (Volts) |
| 1 G | 100 | 0.8 | 40 | 0.06 | 200 |
| 100 M | 50 | 0.4 | 10 | 0.025 | 200 |
| 10 M | 35 | 0.25 | 10 | 0.025 | 200 |
| 1 M | 25 | 0.03 | 2.5 | 0.02 | 200 |
| 100 k | 15 | 0.02 | 2.5 | 0.01 | 100 |
| 10 k | 10 | 0.01 | 2.5 | 0.005 | 32 |
| 1 k | 10 | 0.01 | 2.5 | 0.005 | 10 |
| 100 | 10 | 0.01 | 2.5 | 0.005 | 3.2 |
| 10 | 10 | 0.01 | 2.5 | 0.005 | 1.0 |
| 1 | 10 | 0.01 | 2.5 | 0.005 | 0.32 |
| 0.1 | 100 | 0.1 | 10* | 0.025 | 0.3 |
| Temperature Stability | ± 0.01 °C Over a 1 Year Period | ||||
| Ambient Temperature Range | 23 °C ± 5 °C | ||||
| Ambient Humidity Range | 20% to 50% RH | ||||
| Warranty | 1 year (parts & labour) | ||||
* Stability specifications based on using the shunt as a standard resistor
** 0.1 Ω Power Coefficient < 1 µΩ/Ω
Scanner Specifications
| Operation | 4-terminal |
| Error Contribution | < 20 nV |
| Contact Configuration | Two-coil latching relay |
| Max Carrying/Switching Current | 4/2 A @ < 30 V (DC) |
| Maximum Working/Switching Voltage | 1000/220 V @ < 100 mA (DC) |
| Contact Resistance | < 0.05 Ω |
| Expected Relay Life | 108 operations |
| Insulation Resistance | > 1012 Ω |