1330A Artifact Transfer Standard

Featuring

  • Fully automated artifact calibration of calibrators and DMMs in a single instrument
  • Three standards, one box
  • Primary 1 Ω, 10 kΩ, and 10 V references
Overview

Measurements International’s Artifact Transfer Standard (1330A) is a highly versatile, accurate instrument that meets laboratory requirements for automated artifact calibration (to assign values to internally generated parameters) on calibrators and DMMs. This process is typically performed using a small number of standards at recommended calibration intervals indicated in the calibrator or DMM manufacturer manual. The 1330A is composed of a temperature-controlled instrument enclosure, three reference standards, an external power supply, and a battery backup.

Feature Benefit
Primary 1 Ω, 10 kΩ and 10 V references. All three standards in one temperature-controlled enclosure.
Clear connection for Josephson Volt Comparison of 10 V Zener. Provides highest level traceability.
Manual or automated control. Select from three standards for calibration.
Output connections to the calibrator or DMM on the front panel. Choice of cables available (and supplied) feature direct plug-in for 57XX series calibrators or to the 3458A DMM.
Current standard value is extrapolated from prior calibrations. Enhanced automation and improved accuracy.
Oven temperature monitored by internal PRT. Better artifact performance as the temperature environment is stable and controlled to less than 100 mK.
Calibrated values of the standards entered in the unit. No requirement for set-up.
Front panel display shows certification value. Quick access and reference to information.
Specifications
Artifact Standard TC1 Stability per Year2 Dissipation
10 V 0.1 μV/V/°C 30 Days: 0.25 μV/V
90 Days: 0.75 μV/V
1 Year: 1.5 μV/V
1 Ω 0.1 μΩ/Ω/°C 1 Year: 0.5 μΩ/Ω 10 mW
10 kΩ 0.1 μΩ/Ω/°C 1 Year: 0.5 μΩ/Ω 10 mW
Temperature Stability ± 0.1 °C Over a 1 Year Period
Ambient Temperature Range 23 °C ± 5 °C
Ambient Humidity Range 20% – 90% non-condensing
Storage (with Battery Removed) -50 °C to +50 °C
Isolation to Earth > 1012 Ω
Direct Cable to 57XX Provided
Power 24-hour internal battery3
Warranty 2 years parts & labour

1 Temperature coefficient: 0.2 ppm/°C for ambient temperatures outside ± 5 °C

2 The above specifications are valid after the first 12 months of manufacture due to the natural aging of the internal artifacts.

3 72-hour battery pack (1330A-PS) contained in its own enclosure, plugs into the rear of 1330A